Appendix V H. Viscosity
The dynamic viscosity or viscosity coefficient η is the tangential force per unit surface, known as shearing stress τ and expressed in pascals, necessary to move, parallel to the sliding plane, a layer of liquid of 1 square metre at a rate (v) of 1 metre per second relative to a parallel layer at a distance (x) of 1 metre.
The ratio dv/dx is a speed gradient giving the rate of shear D expressed in reciprocal seconds (s-1), so that η = τ/D.
The unit of dynamic viscosity is the pascal second (Pa⋅s). The most commonly used submultiple is the millipascal second (mPa⋅s).
The kinematic viscosity v, expressed in square metres per second, is obtained by dividing the dynamic viscosity η by the density ρ expressed in kilograms per cubic metre, of the liquid measured at the same temperature, i.e. v = η/ρ. The kinematic viscosity is usually expressed in square millimetres per second.
A capillary viscometer may be used for determining the viscosity of Newtonian liquids and a rotating viscometer for determining the viscosity of Newtonian and non-Newtonian liquids. Other viscometers may be used provided that the accuracy and precision is not less than that obtained with the viscometers described below.
Method I
Apparatus
The apparatus consists of a glass U-tube viscometer (Fig.5H–1) made of clear borosilicate glass and constructed in accordance with the dimensions shown in the figure and in Table 5H–1. The monograph states the size of viscometer to be used.
Method
Fill the viscometer with the liquid being examined through tube L to slightly above the mark G, using a long pipette to minimise wetting the tube above the mark. Place the tube vertically in a water bath and when it has attained the specified temperature, adjust the volume of the liquid so that the bottom of the meniscus settles at the mark G. Adjust the liquid to a point about 5 mm above the mark E. After releasing pressure or suction, measure the time taken for the bottom of the meniscus to fall from the top edge of mark E to the top edge of mark F.
Calculate, as required, either the kinematic viscosity (ν) in square millimetres per second (mm2 s–1) from the expression: ν = Kt,
or the dynamic viscosity (η) in millipascal seconds (mPa s) from the expression: η = Kρt,
where | t | = | time in seconds for the meniscus to fall from E to F, |
ρ | = | mass/volume (g cm–3) obtained by multiplying the relative density, Appendix V G, of the fluid being examined by 0.9982. |
The constant (K) of the instrument is determined using the appropriate European Pharmacopoeia reference liquid for viscometers.
Method II (Capillary viscometer method)
The determination of viscosity using a suitable capillary viscometer is carried out at a temperature of 20 ± 0.1 °C, unless otherwise prescribed. The time required for the level of the liquid to drop from one mark to the other is measured with a stop-watch to the nearest one-fifth of a second. The result is valid only if two consecutive readings do not differ by more than 1 per cent. The average of not fewer than three readings gives the flow time of the liquid to be examined.
Calculate the dynamic viscosity η (2.2.8) in millipascal seconds using the formula:
k | = | constant of the viscometer, expressed in square millimetres per second squared, |
ρ | = | density of the liquid to be examined expressed in milligrams per cubic millimetre, obtained by multiplying its relative density () by 0.9982, |
t | = | flow time, in seconds, of the liquid to be examined. |
The constant k is determined using a suitable viscometer calibration liquid.
To calculate the kinematic viscosity (mm2⋅s-1), use the following formula: v = kt.
The determination may be carried out with an apparatus (Figure 2.2.9.-1) having the specifications described in Table 2.2.9.-11 :
The minimum flow time should be 350 s for size no. 1 and 200 s for all other sizes.
Method Fill the viscometer through tube (L) with a sufficient quantity of the liquid to be examined, previously brought to 20 °C unless otherwise prescribed, to fill bulb (A) but ensuring that the level of liquid in bulb (B) is below the exit to ventilation tube (M). Immerse the viscometer in the bath of water at 20 ± 0.1 °C, unless otherwise prescribed, maintain it in the upright position and allow to stand for not less than 30 min to allow the temperature to reach equilibrium. Close tube (M) and raise the level of the liquid in tube (N) up to a level about 8 mm above mark (E). Keep the liquid at this level by closing tube (N) and opening tube (M). Open tube (N) and measure, with a stop-watch to the nearest one-fifth of a second, the time required for the level of the liquid to drop from mark (E) to (F).
Method III (Rotating viscometer method)
The principle of the method is to measure the force acting on a rotor (torque) when it rotates at a constant angular velocity (rotational speed) in a liquid. Rotating viscometers are used for measuring the viscosity of Newtonian (shear-independent viscosity) or non-Newtonian liquids (shear dependent viscosity or apparent viscosity). Rotating viscometers can be divided in 2 groups, namely absolute and relative viscometers. In absolute viscometers the flow in the measuring geometry is well defined. The measurements result in absolute viscosity values, which can be compared with any other absolute values. In relative viscometers the flow in the measuring geometry is not defined. The measurements result in relative viscosity values, which cannot be compared with absolute values or other relative values if not determined by the same relative viscometer method.
Different measuring systems are available for given viscosity ranges as well as several rotational speeds.
APPARATUS
The following types of instruments are most common.
concentric cylinder viscometers (absolute viscometers)
In the concentric cylinder viscometer (coaxial double cylinder viscometer or simply coaxial cylinder viscometer), the viscosity is determined by placing the liquid in the gap between the inner cylinder and the outer cylinder. Viscosity measurement can be performed by rotating the inner cylinder (Searle type viscometer) or the outer cylinder (Couette type viscometer), as shown in Figures 2.2.10.-1 and 2.2.10.-2, respectively. For laminar flow, the viscosity (or apparent viscosity) η expressed in pascal-seconds is given by the following formula:
M | = | torque in newton-metres acting on the cylinder surface, |
ω | = | angular velocity in radians per second, |
h | = | height of immersion in metres of the inner cylinder in the liquid medium, |
Ri | = | radius in metres of the inner cylinder, |
Ro | = | radius in metres of the outer cylinder, |
k | = | constant of the apparatus, expressed in radians per cubic metre. |
For non-Newtonian liquids it is indispensable to specify the shear stress (τ) or the shear rate (γ) at which the viscosity is measured. Under narrow gap conditions (conditions satisfied in absolute viscometers), there is a proportional relationship between M and τ and also between ω and γ:
where A and B are constants for the instrument and are calculated from the following expressions:
M | = | torque in Newton-metres acting on the cone or cylinder surface, |
ω | = | angular velocity in radians per second, |
Ri | = | radius in metres of the inner cylinder, |
R0 | = | radius in metres of the outer cylinder, |
R | = | radius in metres of the cone, |
h | = | height of immersion in metres of the inner cylinder in the liquid medium, |
α | = | angle in radians between the flat disk and the cone, |
τ | = | shear stress in pascals (Pa), |
γ | = | shear rate in reciprocal seconds (s-1). |
Cone-plate viscometers (absolute viscometers)
In the cone-plate viscometer, the liquid is introduced into the gap between a flat disc and a cone forming a define angle. Viscosity measurement can be performed by rotating the cone or the flat disc, as shown in Figures 2.2.10.-3 and 2.2.10.-4, respectively. For laminar flow, the viscosity (or apparent viscosity) η expressed in pascal-seconds is given by the following formula:
M | = | torque in Newton-metres acting on the flat disc or cone surface, |
ω | = | angular velocity in radians per second, |
α | = | angle in radians between the flat disc and the cone, |
R | = | radius in metres of the cone, |
k | = | constant of the apparatus, expressed in radians per cubic metre. |
Constants A, B of the apparatus (see under concentric cylinder viscometers).
Spindle viscometers (relative viscometers)
In the spindle viscometer, the viscosity is determined by rotating a spindle (for example, cylinder- or disc-shaped, as shown in Figures 2.2.10.-5 and 2.2.10.-6, respectively) immersed in the liquid. Relative values of viscosity (or apparent viscosity) can be directly calculated using conversion factors from the scale reading at a given rotational speed.
In a general way, the constant k of the apparatus may be determined at various speeds of rotation using a certified viscometer calibration liquid. The viscosity η then corresponds to the formula:
METHOD
Measure the viscosity (or apparent viscosity) according to the instructions for the operation of the rotating viscometer. The temperature for measuring the viscosity is indicated in the monograph. For non-Newtonian systems, the monograph indicates the type of viscometer to be used and if absolute viscometers are used the angular velocity or the shear rate at which the measurement is made. If it is impossible to obtain the indicated shear rate exactly, use a shear rate slightly higher and a shear rate slightly lower and interpolate.
With relative viscometers the shear rate is not the same throughout the sample and therefore it cannot be defined. Under these conditions, the viscosity of non-Newtonian liquids determined from the previous formula has a relative character, which depends on the type of spindle and the angular velocity as well as the dimensions of the sample container (Ø = minimum 80 mm) and the depth of immersion of the spindle. The values obtained are comparable only if the method is carried out under experimental conditions that are rigorously the same.
Method IV (Falling Ball and Automatic Rolling Ball Viscometer Methods)
The determination of dynamic viscosity of Newtonian liquids using a falling ball or rolling ball viscometer is performed at 20.0 ± 0.1 °C, unless otherwise prescribed in the monograph. The time required for a test ball to fall or roll in the liquid to be examined from one ring mark or sensor to the other is determined.
method A - falling ball viscometer
Apparatus The falling ball viscometer consists of: a glass tube enclosed in a mantle, which allows precise control of temperature; 6 balls made of glass, nickel-iron or steel with different densities and diameters. The tube is fixed in such a way that the axis is inclined by 10 ± 1° with regard to the vertical. The tube has 2 ring marks that define the distance the ball has to fall. Commercially available apparatuses are supplied with tables giving the constants, the density of the balls and the suitability of the different balls for the expected viscosity range.
Method Fill the clean, dry tube of the viscometer, previously brought to 20.0 ± 0.1 °C, with the liquid to be examined, avoiding bubbles. Choose the ball that is suitable for the viscosity range of the liquid so as to obtain a falling time (run time) not less than 30 s. Place it in the tube, close the tube and maintain the solution at 20.0 ± 0.1 °C for at least 15 min. Let the ball run through the liquid between the 2 ring marks once without measurement. Let it run again and measure with a stop-watch, to at least the nearest 1/5 s, the time required for the ball to fall from the upper to the lower ring mark. Repeat the test run at least 3 times to obtain a minimum of 4 run times. The result is only valid if 2 consecutive run times do not differ by more than 1.5 per cent.
Use the mean of the run times to calculate the dynamic viscosity (η) in millipascal seconds using the formula:
k | = | constant, in square millimetres per second squared; |
ρ1 | = | density of the ball used, in grams per cubic centimetre; |
ρ2 | = | density of the liquid to be examined, in grams per cubic centimetre, obtained by multiplying its relative density () by 0.9982; |
t | = | mean of the run times of the ball, in seconds. |
method B – AUTOMATIC ROLLING BALL VISCOMETER
Apparatus The automatic rolling ball viscometer consists of: several capillaries made of glass or other suitable materials with different diameters enclosed in a thermostatically controlled block, which allows precise control of the temperature; balls made of stainless steel (optionally coated) or other suitable materials; a motor drive that positions the capillary at an inclination angle from 10.0 ± 0.2° to 80.0 ± 0.2° with regard to the vertical. The apparatus has at least 2 sensors to measure the run time of the ball over a predefined distance. Commercially available apparatuses are supplied with tables giving the constants, the density of the balls and the suitability of the different capillaries for the expected viscosity range. Temperature control, control of the inclination angle, run-time determination, repetitions of runs and calculations of the mean and relative standard deviation are performed by the apparatus software.
Method Set the instrument software to perform a measurement with at least 4 run times (2 forward/backward cycles) and a maximum relative standard deviation of 0.5 per cent. Choose the capillary, the ball and the inclination angle that are suitable for the expected viscosity range of the liquid to be examined so as to obtain a rolling time (run time) not less than 20 s over the 100 mm distance or a proportional time over other distances. If no digital density meter is connected to the viscometer, ensure that the density value of the liquid to be examined has been specified in the apparatus software. Fill the clean, dry capillary of the viscometer with the liquid to be examined, avoiding bubbles. Start the measurement immediately after filling the capillary.
The instrument calculates automatically the dynamic viscosity (η) in millipascal seconds and the kinematic viscosity (v) in square millimetres per second.
Additional points for monographs of the British Pharmacopoeia
The apparatus and methods described in Methods I and II are in agreement in all essentials with International Standards ISO 3104-1994 and 3105-1994 (Methods for the determination of the viscosity of liquids).