Appendix IX K. Gas Detector Tubes
Gas detector tubes are cylindrical, sealed tubes consisting of an inert transparent material and are constructed to allow the passage of gas. They contain reagents adsorbed onto inert substrates that are suitable for the visualisation of the substance to be detected and, if necessary, they also contain preliminary layers and/or adsorbent filters to eliminate substances that interfere with the substance to be detected. The layer of indicator contains either a single reagent for the detection of a given impurity or several reagents for the detection of several substances (monolayer tube or multilayer tube).
The test is carried out by passing the required volume of the gas to be examined through the indicator tube. The length of the coloured layer or the intensity of a colour change on a graduated scale gives an indication of the impurities present.
The calibration of the detector tubes is verified according to the manufacturer′s instructions.
Operating conditions Examine according to the manufacturer′s instructions or proceed as follows.
The gas supply is connected to a suitable pressure regulator and needle valve. Connect the flexible tubing fitted with a Y-piece to the valve and adjust the flow of gas to be examined to purge the tubing in order to obtain an appropriate flow (Figure 2.1.6.-1). Prepare the indicator tube and fit to the metering pump, following the manufacturer′s instructions. Connect the open end of the indicator tube to the short leg of the tubing and operate the pump by the appropriate number of strokes to pass a suitable volume of gas to be examined through the tube. Read the value corresponding to the length of the coloured layer or the intensity of the colour on the graduated scale. If a negative result is achieved, indicator tubes can be verified with a calibration gas containing the appropriate impurity.
In view of the wide variety of available compressor oils, it is necessary to verify the reactivity of the oil detector tubes for the oil used. Information on the reactivity for various oils is given in the leaflet supplied with the tube. If the oil used is not cited in the leaflet, the tube manufacturer must verify the reactivity and if necessary provide a tube specific for this oil.
Arsine detector tube
Sealed glass tube containing adsorbent filters and suitable supports for the gold salt or other appropriate indicator. The minimum value indicated is 0.25 ppm or less, with a relative standard deviation of at most 20 per cent.
Carbon dioxide detector tube
Sealed glass tube containing adsorbent filters and suitable supports for hydrazine and crystal violet indicators. The minimum value indicated is 100 ppm with a relative standard deviation of at most 15 per cent.
Carbon monoxide detector tube
Sealed glass tube containing adsorbent filters and suitable supports for di-iodine pentoxide, selenium dioxide and fuming sulfuric acid indicators. The minimum value indicated is 5 ppm or less, with a relative standard deviation of at most 15 per cent.
Hydrogen sulfide detector tube
Sealed glass tube containing adsorbent filters and suitable supports for an appropriate lead salt indicator. The minimum value indicated is 0.2 ppm or less, with a relative standard deviation of at most 10 per cent.
Nitrogen monoxide and nitrogen dioxide detector tube
Sealed glass tube containing adsorbent filters and suitable supports for an oxidising layer (Cr(VI) salt) and the diphenylbenzidine indicator. The minimum value indicated is 0.5 ppm with a relative standard deviation of at most 15 per cent.
Oil detector tube
Sealed glass tube containing adsorbent filters and suitable supports for the sulfuric acid indicator. The minimum value indicated is 0.1 mg/m3 with a relative standard deviation of at most 30 per cent.
Phosphine detector tube
Sealed glass tube containing adsorbent filters and suitable supports for the gold salt or other appropriate indicator. The minimum value indicated is 0.2 ppm or less, with a relative standard deviation of at most 20 per cent.
Sulfur dioxide detector tube
Sealed glass tube containing adsorbent filters and suitable supports for the iodine and starch indicator. The minimum value indicated is 0.5 ppm with a relative standard deviation of at most 15 per cent.
Water vapour detector tube
Sealed glass tube containing adsorbent filters and suitable supports for the magnesium perchlorate indicator. The minimum value indicated is 67 ppm or less, with a relative standard deviation of at most 20 per cent.
1. Gas supply | 5. Indicator tube | |
2. Pressure regulator | 6. Indicator tube pump | |
3. Needle valve | 7. End open to atmosphere | |
4. Y-piece |